Presage Arterial Pressure Waveform and Pulse Rate Model Card


THE FOLLOWING INFORMATION, AND PRESAGE'S SDK AND APP, ARE OFFERED FOR GENERAL WELLNESS AND INFORMATIONAL PURPOSES ONLY. NEITHER HAVE BEEN CLEARED BY THE FDA AND NEITHER MAY BE USED FOR MEDICAL DIAGNOSIS OR TREATMENT.

1. Model Details

Basic info: Presage vitals by video analysis generates a relative arterial pressure waveform and pulse rate from a video of a subject. This card covers 10.6-second pulse rate (vs EKG) and 60-second waveform analysis (vs BP).

Organization: Presage Technologies

Model date: 2026-09-29

Model version: fed9fe66

Model type: A proprietary computer vision and signal processing pipeline extracts a relative arterial pressure waveform and pulse rate from video of a subject. Pulse rate is compared against EKG at 10.6s windows. The waveform is z-score normalized and compared against the Biopac BP reference at 60s windows.

License: The algorithm is currently proprietary, and licenses are granted with predefined agreement.

Contact: Questions can be sent to: support@presagetech.com


2. Intended Use

Model Uses

This model was intended for use in the analysis and non-diagnostic utility of arterial pressure waveform morphology and pulse rate. Requires a stationary device and video of a subject acquired at 25+ fps. Pulse rate range: 40-180 BPM.

Out-of-Scope Uses

As noted above. Not intended for diagnostic blood pressure measurement. Does not provide absolute systolic, diastolic, or MAP values in mmHg. Not an arrhythmia detection or monitoring device.


3. Validation

Reference Standard: Biopac MP160 3-lead ECG (for pulse rate comparison at 10.6s windows) and Biopac continuous blood pressure monitor (for waveform correlation at 60s windows). Waveforms z-score normalized before comparison.

Comparison Methodology: Arterial Pressure Waveform measurements from the camera-based system were compared against time-aligned reference measurements. Ground truth signals were checked for quality using labeled signal annotations; segments with poor signal quality were excluded from analysis.


4. Data Demographics

Category Distribution
Total 118 subjects, 418 videos
Camera (videos) e-con See3CAM CU27: 114, Logitech C920: 113, Samsung S24 Rear 24mm (tripod): 101, Samsung S24 Front 27mm (handheld): 90
Sex (subjects) Female: 63, Male: 55
Age Group (subjects) 18-25: 33, 26-35: 39, 36-45: 18, 46-55: 14, 56-65: 11, 65+: 2
Fitzpatrick (subjects) Type 1: 21, Type 2: 12, Type 3: 13, Type 4: 24, Type 5: 33, Type 6: 15
BMI (subjects) Mean: 27.9, Range: 16.5–54.3, N=118
Lighting (videos) Ring Light: 133, Lamp: 132, Window: 99

Reference standard: Biopac EKG and continuous blood pressure monitor.


5. Data Provenance

Reference Instrumentation: Biopac research-grade physiological sensors: 3-lead ECG (for pulse rate ground truth), continuous blood pressure monitor (for waveform correlation ground truth).

Camera Devices Tested: Samsung S24 Rear 24mm (tripod), Samsung S24 Front 27mm (handheld), e-con See3CAM CU27 (USB), Logitech C920 (tripod).

Average Camera Distance: Samsung S24/e-con/Logitech: 26", Samsung S24 handheld: 10"

Data Handling: All subject data is de-identified. Derived metrics and anonymized identifiers are retained. Data is securely stored with access restricted to trained researchers.


6. Factors

The arterial pressure model requires face detection and high-quality photoplethysmographic signal extraction.


These factors can affect model performance:


Lighting Conditions Tested:


Other factors:


7. Metrics

(at 80% Return Rate, Confidence >= 97)

  1. MAE: 0.36 BPM
  2. RMSE: 0.64 BPM

8. Quantitative Analysis

Computed vs Ground Truth at Confidence Thresholds

Bland-Altman Plot