Presage Arterial Pressure Waveform and Pulse Rate Model Card


THE FOLLOWING INFORMATION, AND PRESAGE'S SDK AND APP, ARE OFFERED FOR GENERAL WELLNESS AND INFORMATIONAL PURPOSES ONLY. NEITHER HAVE BEEN CLEARED BY THE FDA AND NEITHER MAY BE USED FOR MEDICAL DIAGNOSIS OR TREATMENT.

1. Model Details

Basic info: Presage vitals by video analysis generates a relative arterial pressure waveform and pulse rate from a video of a subject. This card covers 10.6-second pulse rate (vs EKG) and 60-second waveform analysis (vs BP).

Organization: Presage Technologies

Model date: 2026-09-29

Model version: fed9fe66

Model type: A proprietary computer vision and signal processing pipeline extracts a relative arterial pressure waveform and pulse rate from video of a subject. Pulse rate is compared against EKG at 10.6s windows. The waveform is z-score normalized and compared against the Biopac BP reference at 60s windows.

License: The algorithm is currently proprietary, and licenses are granted with predefined agreement.

Contact: Questions can be sent to: support@presagetech.com


2. Intended Use

Model Uses

This model was intended for use in the analysis and non-diagnostic utility of arterial pressure waveform morphology and pulse rate. Requires a stationary device and video of a subject acquired at 25+ fps. Pulse rate range: 40-180 BPM.

Out-of-Scope Uses

As noted above. Not intended for diagnostic blood pressure measurement. Does not provide absolute systolic, diastolic, or MAP values in mmHg. Not an arrhythmia detection or monitoring device.


3. Validation

Reference Standard: Biopac MP160 3-lead ECG (for pulse rate comparison at 10.6s windows) and Biopac continuous blood pressure monitor (for waveform correlation at 60s windows). Waveforms z-score normalized before comparison.

Comparison Methodology: Arterial Pressure Waveform measurements from the camera-based system were compared against time-aligned reference measurements. Ground truth signals were checked for quality using labeled signal annotations; segments with poor signal quality were excluded from analysis.


4. Data Demographics

Category Distribution
Total 118 subjects, 418 videos
Camera (videos) e-con See3CAM CU27: 114, Logitech C920: 113, Samsung S24 Rear 24mm (tripod): 101, Samsung S24 Front 27mm (handheld): 90
Sex (subjects) Female: 63, Male: 55
Age Group (subjects) 18-25: 33, 26-35: 39, 36-45: 18, 46-55: 14, 56-65: 11, 65+: 2
Fitzpatrick (subjects) Type 1: 21, Type 2: 12, Type 3: 13, Type 4: 24, Type 5: 33, Type 6: 15
BMI (subjects) Mean: 27.9, Range: 16.5–54.3, N=118
Lighting (videos) Ring Light: 133, Lamp: 132, Window: 99

Reference standard: Biopac EKG and continuous blood pressure monitor.


5. Data Provenance

Reference Instrumentation: Biopac research-grade physiological sensors: 3-lead ECG (for pulse rate ground truth), continuous blood pressure monitor (for waveform correlation ground truth).

Camera Devices Tested: Samsung S24 Rear 24mm (tripod), Samsung S24 Front 27mm (handheld), e-con See3CAM CU27 (USB), Logitech C920 (tripod).

Average Camera Distance: Samsung S24/e-con/Logitech: 26", Samsung S24 handheld: 10"

Data Handling: All subject data is de-identified. Derived metrics and anonymized identifiers are retained. Data is securely stored with access restricted to trained researchers.


6. Factors

The arterial pressure model requires face detection and high-quality photoplethysmographic signal extraction.


These factors can affect model performance:


Lighting Conditions Tested:


Other factors:


7. Metrics

(at 80% Return Rate, Confidence >= 97)

  1. MAE: 0.36 BPM
  2. RMSE: 0.64 BPM

8. Quantitative Analysis

Computed vs Ground Truth at Confidence Thresholds

Bland-Altman Plot

Confidence Lookup Table

Confidence >= MAE (BPM) RMSE (BPM) Pearson r Return Rate (%) N (samples)
0 0.92 2.91 0.975 100.0 25117
10 0.89 2.76 0.977 99.6 25024
20 0.86 2.66 0.979 99.4 24954
30 0.81 2.46 0.982 98.8 24815
40 0.77 2.31 0.984 98.2 24663
50 0.70 2.08 0.987 97.3 24436
60 0.63 1.77 0.991 96.0 24104
65 0.58 1.58 0.992 95.1 23893
70 0.55 1.40 0.994 94.3 23687
75 0.51 1.23 0.995 93.4 23452
80 0.47 1.01 0.997 92.3 23186
85 0.45 0.89 0.998 91.0 22867
90 0.42 0.77 0.998 89.4 22446
95 0.38 0.67 0.999 85.1 21372
96 0.37 0.66 0.999 83.0 20842
97 0.36 0.64 0.999 80.0 20075
98 0.35 0.61 0.999 74.4 18679
99 0.33 0.59 0.999 63.8 16031

Performance

(at 80% Return Rate, Confidence >= 97)

By Camera Type

Camera Type N (samples) Return Rate (%) MAE (BPM) RMSE (BPM) Pearson r
Samsung S24 Rear 24mm (tripod) 6016 85.1 0.43 0.91 0.998
Samsung S24 Front 27mm (handheld) 1261 63.5 0.37 0.53 0.999
e-con See3CAM CU27 6259 78.3 0.35 0.48 0.999
Logitech C920 6539 81.0 0.32 0.46 0.999

By Fitzpatrick Skin Type

Fitzpatrick N (samples) Return Rate (%) MAE (BPM) RMSE (BPM) Pearson r
Type I 4036 93.3 0.33 0.53 0.999
Type II 2392 89.4 0.27 0.38 1.000
Type III 2653 93.1 0.49 1.13 0.997
Type IV 4417 87.1 0.33 0.49 0.999
Type V 5171 73.3 0.36 0.52 0.999
Type VI 1406 44.7 0.48 0.77 0.998
Fitzpatrick Type VI: Breakdown by Lighting Condition
Ring Light 539 54.7 0.37 0.49 0.997
Window 311 32.5 0.52 0.74 0.999
Lamp 419 39.5 0.60 1.07 0.986

Note: Types I-V are composites across all lighting conditions (Ring Light, Window, Lamp).

By Sex

Sex N (samples) Return Rate (%) MAE (BPM) RMSE (BPM) Pearson r
Male 10093 82.7 0.35 0.69 0.999
Female 9982 77.3 0.38 0.58 0.999

By Age Group

Age Group N (samples) Return Rate (%) MAE (BPM) RMSE (BPM) Pearson r
18-25 5586 82.2 0.46 0.89 0.997
26-35 6563 80.5 0.33 0.46 0.999
36-45 2765 73.2 0.38 0.62 0.999
46-55 2733 85.1 0.28 0.48 0.999
56-65 2191 85.8 0.27 0.37 0.999
65+ 61 15.4 0.49 0.73 0.999

By Lighting Type

Lighting N (samples) Return Rate (%) MAE (BPM) RMSE (BPM) Pearson r
Lamp 6110 77.3 0.42 0.85 0.997
Ring Light 6871 83.6 0.34 0.50 0.999
Window 4112 72.2 0.36 0.57 0.999

Confidence vs Waveform Correlation

Confidence >= Waveform Pearson r Return Rate (%) N (videos)
0 0.681 100.0 375
10 0.681 100.0 375
20 0.683 99.7 374
30 0.683 99.7 374
40 0.689 98.4 369
50 0.693 97.3 365
60 0.699 95.5 358
65 0.702 94.4 354
70 0.704 93.1 349
75 0.707 91.5 343
80 0.707 90.1 338
85 0.714 88.8 333
90 0.721 84.3 316
95 0.729 77.6 291
96 0.733 76.0 285
97 0.736 72.8 273
98 0.734 67.2 252
99 0.731 56.5 212

Waveform Example


9. Fairness & Equity

Bias Assessment Methodology: Performance is stratified by Fitzpatrick skin type (I-VI), sex, camera type, and age group. Per-group metrics and Confidence averages are reported in the Quantitative Analysis tables above.


10. Ethical Considerations

As a remote sensing device, the risks posed to the subjects in the trial are minimal, including the association of each subject with corresponding biometric data. Mitigation of these risks include de-identifying all subject data, including videos, prior to saving it. Additionally, all data is securely stored with access to a select number of trained researchers.

The model is not intended for human life-critical decisions, diagnostics or prognostication.

11. Limitations and Tradeoffs